pH Meters
Innovation in pH Measurement
|
|
MP230 GLP Research pH/mV/°C Meter
|
Specifications |
pH-range: |
-2.00 to 19.999 |
pH-resolution: |
0.001 |
pH-relative accuracy:
|
± 0.001 |
mV-range: |
-1999.9 to 1999.9
|
mV-resolution: |
0.1 |
mV-relative accuracy:
|
± 0.1 |
Temperature range °C:
|
-5.0 to 130.0 |
Temperature
resolution °C: |
0.1 |
Temperature accuracy
°C: |
± 0.2 |
Display: |
Custom LCD |
Power supply: |
9V DC from supplied
transformer |
|
|
|
Description
Equipped with a host of features, the MP230 is ideal for the most
demanding applications in research & development or quality control
laboratories where accuracy and flexibility are both critical issues.
|
Features and Benefits
The MP230 has a number of additional features specifically designed for
GLP. For critical quality control applications an alarm can be set which
activates when sample values fall outside pre-set high and low limits.
The MP230´s memory allows the last 10 calibrations to be printed out
whilst the user can select any four-digit sample identification number.
The calibration reminder has an extended range up to 9 hours.
The MP230´s programmable electrode offset and slope limits ensures that
the electrode can only be used when performing within acceptable
customer criteria.
- 3 point pH calibration
- -2 to 19.999 pH range
- Selectable display resolution 0.001/0.01/0.1
- Autocalibration with choice of 11 buffers
- Auto buffer prompting
- RS232 data output
- Additional GLP (Good Laboratory Practice) features
|