Mettler Toledo - Navigeringsmeny |
|
pH Meters
Innovation in ISE Measurement
|
|
MA235 Advanced pH/Ion/°C Meter
|
Specifications |
pH-range: |
-2.000 to 19.999 |
pH-relative accuracy:
|
± 0.001 |
mV-range: |
-1999.9 to 1999.9 |
mV-relative accuracy:
|
± 0.1 |
Temperature range °C:
|
-30 to 130 |
Temperature accuracy °C:
|
± 0.5 |
Concentration range: |
1.00E-9 to 9.99E+9 |
Conc. accuracy: |
± 0.5 % |
Display: |
Graphics LCD backlit |
Power supply: |
9V DC from supplied
transformer |
|
|
|
Description
The MA235 is the flagship of the Quattro range combining advanced level ion
concentration and pH measurement in one unit.
The MA235 has all the features of the MP230 advanced pH meter but takes GLP
to a new level. The MA235 can be set-up to enforce calibration after either
a preset time interval or number of samples to ensure compliance with
customer calibration regimes.
Dual electrode inputs and split screen display ensure the two completely
seperate measurement systems are seamlessly combined in a single benchtop
meter.
|
Features and Benefits
An electrode switch box can be added to either channel allowing connection
for up to five electrodes, making it unnecessary to change electrodes when
switching sampling methods. Calibration data and electrode identification
number (e.g. serial number) for each electrode is stored when setting up
each method. A further aid in set-up is the two-way RS232 port which allows
the MA235 to be set-up remotely from a PC.
The five point calibration ensures a high level of accuracy even at the
extremes of concentration measurement whilst a full range of incremental
techniques, in addition to direct ion measurement, ensures all ISE
applications are covered.
- 5 point pH and concentration calibration
- -2 to 19.999 pH range
- Selectable display resolution 0.001/0.01/0.1
- Autocalibration with choice of 13 buffers
- Auto buffer prompting
- Up to three significant figures in concentration
- 100 point internal data logging
- Two way RS232 communications
- Dual electrode inputs
- Incremental techniques
- Temperature compensated ISE
- Electrode selector box (optional)
- Timed endpoint
- Ten method storage (each channel)
- 5 language menu system
- Large clear backlit LCD display
- Comprehensive GLP (Good Laboratory Practice) features
|
|
|